IFU AND STERILIZATION

Instructions and directions for use.

Every Bontempi instrument is manufactured and inspected with the utmost care and is intended exclusively for professionally qualified healthcare practitioners and for the use for which it was designed and built.

First Use

Cleaning

The instrument is supplied in a “NON-STERILE” condition and lubricated with a special protective oil in the joints, hinged parts, and threaded components. Therefore, prior to use, instruments must be thoroughly washed using neutral detergents and/or degreasers suitable for steel, so as not to cause unwanted complications to the instrument (red stains, stiffening of hinged parts, etc.).

After using the instruments, rinse them immediately to remove all residues using a stiff brush with plastic bristles (NO METAL), or, to avoid direct contact, immerse the instruments in a DISINFECTANT-DETERGENT solution for AT LEAST 10 MINUTES.

Avoid ultrasonic bath cleaning for all instruments with cutting edges (scissors, scalpels, bone rongeurs, chisels) and those with tungsten carbide inserts, as they may chip, or the brazed joints of the inserts may detach or become damaged.
NEVER PLACE INSTRUMENTS MADE OF DIFFERENT MATERIALS IN THE SAME CLEANING CYCLE.

After cleaning, ensure that the instruments are completely dry and lubricate moving parts, hinges, and joints with products specifically designed for surgical instruments (NO INDUSTRIAL LUBRICANTS OR OILS).

Sterilization

Warnings

Sterilize instruments only after thorough cleaning and lubrication. It is also important to follow the instructions for use provided by the manufacturer of the sterilization equipment.
STEAM STERILIZATION (AUTOCLAVE): Do not exceed 134°C; hinged or jointed instruments must never be locked shut (metal against metal), otherwise cracks or fractures may occur near the joints or hinges due to thermal expansion.
DRY HEAT STERILIZERS OR OVENS: Given the high temperatures reached during sterilization, premature loss of sharpness and slight browning of the instrument may occur, particularly in cutting instruments (scissors, bone rongeurs, chisels).
COLD STERILANTS: We recommend using solutions with an alkaline pH and strictly adhering to the prescribed immersion times of the instruments in the sterilizing solution.

Worn, damaged, or oxidized instruments must not be used, as they are no longer able to perform their intended function. Please note that certain types of damage (corrosion, rust spots, and the like) can transfer to intact instruments.

Failure to comply with these fundamental rules may cause irreparable damage to the instrument and will void any right to warranty replacement or recourse against Bontempi - Bmed Srl.

The instruments supplied by Bontempi - Bmed Srl are subject to Directive 93/42/EEC and must therefore be disposed of in accordance with applicable regional regulations.